VARIOKLAV Classic Laboratory Autoclaves

Email Print

Complete solutions for routine sterilisation applications

VI_classicline_003_k

HP Medizintechnik VARIOKLAV® steam sterilisers are designed and manufactured to exacting quality standards for safe, reproducible performance. They are equipped with everything that’s required for sterilisation tasks in the lab.


Easy, efficient, reliable

Our VARIOKLAV Classic autoclaves are easy to install, use and clean, require little maintenance, and occupy a minimum of floor space.
These robust, cost-effective systems provide 35, 75 or 135 liter chamber capacities and are available with exhaust air filtration for pathogenic waste.

 

VARIOKLAV Classic autoclaves support the rapid, efficient sterilisation of instruments, liquids and waste across a range of environments:
• Education
• Academic research
• Food and beverage quality control
• Climate chamber for environmental
testing
• Research institutes

There are floor and tabletop models available.

 

SPA100_c_300

Simple, user-friendly microprocessor control

Simple, user-friendly microprocessor control VARIOKLAV Classic autoclaves ar e equipped with an easy-to-use control featuring intuitive pictograms and clear displays of process parameters.

• Easy, one-touch recall of three sterilisation programs (two pre-set, one user-configurable) for liquids, equipment/solid materials and waste


• Audible and visual alerts upon successful program completion

S4_4_b_b_300
SPA108_db_300



Convenience and simplicity in steam sterilisation

Fully equipped and ready for work
VARIOKLAV Classic autoclaves are supplied with everything you need for your sterilisation applications:
• Media temperature probe for precise control of liquid sterilisation times and temperature
• Tamper-proof thermolock prevents retarded boiling when sterilizing liquids
S18_2_c_300 • Condensate collector for exhaust steam
• Two high-capacity, stainless steel sterilisation baskets

VARIOKLAV Classic autoclaves are delivered ready to start work when you are. Installation is straightforward, because no connection to a feed water supply, tap water nor a waste water connection is required. Simply add demineralized water to the chamber and initiate your program. All phases of the sterilisation process run automatically, allowing safe, unattended operation.

Easy and comfortable to use
• Intuitive microprocessor controls automate every aspect of the sterilisation process
• Security key switch saves program settings
• Side-opening, quick-release lid minimizes risk from rising steam
• Low deck height for easier loading and unloading
• Short heating times
• No heating elements in sterilisation chamber for maximized useful capacity (floor model)
• Continuous venting to the principle of gravity prevents the appearance of air pockets

 

Safer by design
VARIOKLAV Classic autoclaves are designed with utmost operator safety in mind. They comply with the DIN 58951-2 standard, and are certified to category II or III under the European Pressure Equipment Directive.


 

Ideal for use with thermally unstable media, the
reference temperature probe enables precise
control of sterilisation times and temperature


logo_med_com_300S5_2_b_300

A heat-resistant collector
condenses the exhaust steam

SPA113_b_300 Safety assured
The lid locking mechanism
engages six massive
stainless steel bolts.
Leakage or opening under
pressure is impossible

shutterstock_62756968_300




Safe sterilisation of hazardous materials

Our VARIOKLAV Classic 400 FA and 500 FA autoclaves are additionally equipped to provide exhaust-air filtration and condensate sterilisation.

• Emission-free sterilisation of hazardous biological materials, including pathogens, infectious agents and genetically engineered samples
• Both the external atmosphere and exhaust steam collection system are protected from contamination



 

VVARIOKLAV Classic 400 FA and 500 FA autoclaves
with Exhaust Air Filtration

  • • Aerosols contained in exhaust air are retained by
    an easy-to-exchange filter prior to sterilisation
  • • Simultaneous sterilisation of chamber contents,
    collected condensate and filter housing
  • • Sterile disposal of potentially infectious laborator
    waste
  • • Steam is generated in the chamber,
    so no additional condensate heating is required
S5_3_b_300



Ordering information and technical specifications

VARIOKLAV Classic Laboratory Autoclaves – Floor and tabletop models

Floor

Model Order No. Chamber
Net Volume
(l)
Interior
dimensions
D/L (mm)
Exterior
dimensions
W/L/H (mm)
Operating
temperature
(°C)
Max. operating
pressure
(bar)
Connected load/
heat output
(kW)
Electrical
supply
Weight
(kg)
VARIOKLAV Classic 300 50085003 35/38 300/500  500/520/865  98 - 134 2,5 2,4/2,2 230V/50Hz 80
VARIOKLAV Classic 400 50109674 75/80 400/600  600/620/965 3,6/3,4 120
VARIOKLAV Classic 400 FA
with Exhaust Air Filtration
50109686 75/80 400/600  600/620/965 3,6/3,4 120
VARIOKLAV Classic 500 50109685 135/142  500/700  700/720/1.065 6,5/6,0 3N/400V/50Hz 180
VARIOKLAV Classic 500 FA
with Exhaust Air Filtration
50109687  135/142 500/700  700/720/1.065 6,5/6,0 180

Tabletop

Model Order No. Chamber
Net Volume
(l)
Interior
dimensions
D/L (mm)
Exterior
dimensions
W/L/H (mm)
Operating
temperature
(°C)
Max. operating
pressure
(bar)
Connected load/
heat output
(kW)
Electrical
supply
Weight
(kg)
VARIOKLAV Classic 300 50085004 35/38 300/500 500/520/865 98 - 134 2,5 3,2/3,0 230V/50Hz 80
VARIOKLAV Classic 400 50085005 75/80 400/600 600/620/965 5,3/5,0 3N/400V/50Hz 120
VARIOKLAV Classic 400 FA
with Exhaust Air Filtration
50085006 75/80 400/600 600/620/965 5,3/5,0 120
VARIOKLAV Classic 500 50085001 135/142 500/700 700/720/1.065 6,5/6,0 180
VARIOKLAV Classic 500 FA
with Exhaust Air Filtration
0085002 135/142 500/700 700/720/1.065 6,5/6,0 180